Notebook Color Doppler Ultrasonic Scanner

Model: XF3700-P3
Place of Origin: Sichuan,China (Mainland)
Brand Name: XianFeng
3D doppler scan: 3D Laptop color doppler
laptop ultrasound: leg ultrasound
venous doppler ultrasound: ultrasound scanner machine
doppler ultrasonic scanner: Power Doppler
b mode ultrasound: abdominal ultrasound
heart doppler scan: ultrasonic machine

inquiry

skype basketAdd to Basket  Edit

Share |

Product Description

Payment & Shipping Terms Supply Capacity
Unit Price:1.0 USDProduction Capacity:3000
Trade Term:FOB,CFR,CIF,CIP,CPT,FCAPacking:Package...
Payment Terms:Paypal, L/C, T/T, WU, Money GramDelivery Date:Within 2 Days
Min. Order:1 Set/Sets
Means of Transport:Ocean, Air, Land

This is a notebook full digital color ultrasound diagnosis system, is a research and development of XIANFENG team combined with market demand in the XF7800 type main based on the development and design new style,thin and light, good carrying, appearance is compact, powerful, with types of probe can support more, image processing and measuring package software is rich, full blood flow image clarity, auxiliary functions such as adding PW envelope and three imaging modes and practical features such as real-time synchronization.It is widely used in the examination of abdomen, gynecology, obstetrics, urology, blood vessels, heart, bone and small organs, especially in the clinical examination of large medical institutions and primary medical system.Is a cost-effective color doppler ultrasound diagnostic instrument.

                             notebook full digital color ultrasound diagnosis system

Main Functions:

Having a full digital beam forming technology

Scanning mode: Convex array,Trans vagina, high-frequency linear array, Micro-convex;

Dynamic range: 0~120dB adjustable;

Display mode: B,B/B,M,B/M,CFM,CMF/B,PDI,B/PW, total eight mode;

Application mode: abdomen, gynecology, obstetrics, superficial organ, urologist, heart and user defined model 1-4, total ten models;

Image mode: digital beam forming, tissue harmonic imaging;

Acoustic output: Mechanical index and thermal index real-time display;

Acoustic power:Step is adjustable, real-time display;

Gray scale: 256 scales;

Depth display: ≥250mm;

B/D dual-purpose: linear array: B/PWD; convex array: B/PWD;

Pseudo color processing: 16 kinds of pseudo color encoding can optional;

Gain adjusts: 8 segments TGC, B/M/D/C is independently adjustable; TGC curve can show and hide automatically;

Image magnification: picture in picture zoom in and zoom part function;

Self-motion optimize function: Built-in multiple check type, according to different inspection organs, preset best image check condition, reduce the adjusting operation keys;

One-click optimization function: preset several parameters adjusting focus on a Key, a key to realize image fast optimization;

Measurement and calculation: B mode routine measurement:

Distance, circumference, area, volume, angle, ratio, and stenos rate.

M mode routine measurement: Heart rate, time, distance, speed, ratio, etc.

Gynecology measurement: Uterus, cervix, endometrial, ovary, follicular.

Obstetrics measurement: 

EGA, ETD, fetal weight estimation, AFI index, OB report (including OB tables).

Cardiology measurement: LV measurement.

Urology measurement: 

Prostate volume, displacement volume, bladder capacity, and residual urine output.

PW measurements: Time, speed, Heart Rate, RI, PI, etc.

Other measurement: Slice volume measurement, hip joint angle measurement.

Image storage: Image storage, video storage, cine loop, disk storage capacity≥160G;

Patient data: Medical record management, report inquiry and printing, image video output( HDD ,USB,Optional DVD-RW),built-in ultrasound workstation;

Reporting system: automatic report generation system, and can be full screen characters in both Chinese and English editor;

Output interface: USB,DICOM interface;

                                   notebook full digital color ultrasound diagnosis system

Main Technical Indexes:

The performance requirements of gray-scale imaging mode

The color ultrasonic at the gray-scale imaging performance mode should comply with the provisions of the table 1.1

Table 1.1 At the Gray-scale imaging mode the performance of the probe

performance   indexes

probe type and nominal   frequency

2.0≤f<4.0

2.0≤f<5.0

5.0≤f<8.0

5.0≤f<12.0

a) probe type and model

Micro-convex

Convex array

(type   TC60A)

Cavity

(type   TC10A)

Linear array

(type TL40A)

b) nominal frequency (MHz)

5

3.5

6.5

7.5

c) Scan depth(mm)

≥90

≥160

≥40

≥50

d) Lateral resolution (mm)

≤3(depth≤50)

≤4(80<depth≤60)

≤3(depth≤80)

≤4(80<depth≤130)

≤2(depth≤30)

≤2(depth≤40)

e) Axial resolution (mm)

≤2(depth≤50)

≤2(depth≤80)

≤3(80<depth≤130)

≤1(depth≤40)

≤1(depth≤50)

f) Blind area (mm)

≤7

≤5

≤4

≤3

g) Transverse geometry   precision (%)

≤20

≤15

≤10

≤10

h) Longitudinal geometric location   accuracy (%)

≤10

≤10

≤5

≤5

i) Slice thickness (mm)

≤5

≤5

≤5

≤5

j) Perimeter and area    measured deviation (%)

≤±20

≤±20

≤±20

≤±20

k)  M mode time   display error (%)

≤±10

≤±10

≤±10

≤±10

The performance requirements of color Doppler imaging mode:

a). The color Doppler imaging mode should comply with the provisions of thetable 1.2;

b). Color blood flow image should be essentially coincident with the gray-scale image of pipe's;

c). Blood flow direction should be able to correctly identify, no aliasing phenomenon;

Table1.2 at the color blood flow imaging mode the performance of the probe

Doppler model

Micro

Convex array

Cavity

Linear array

Investigation depth at Color   blood flow model

≥90mm

≥100mm

≥40mm

≥50mm

Investigation depth at   Doppler spectrum model

≥90mm

≥100mm

≥40mm

≥50mm

Blood flow speed reading   error

≤±15%

The performance requirements of Doppler spectrum mode:

a). The color ultrasonic at the color Doppler spectrum mode should comply with the provisions of the;

b). Blood flow speed reading error should comply with the provisions of the table2.3;

c). Pulse wave Doppler mode sampling area cursor position should be accurate;

Displayer: 15 inch LCD color display

Running hours: ≥8h;

Input power: ≤300VA;

Host weight: about 6 kg;

Host appearance size: 370 ×382×90(length × width × height) (mm3).